Vibration Tests
- Mil-Std-883 Methods 2005 and 2007
Vibration Tests
are tests performed to determine the effects of
mechanical vibration within a specified frequency range on semiconductor devices.
There are two military standards that are widely used for this purpose:
1) the
Vibration Fatigue Test,
as defined by Mil-Std-993 Method 2005; and
2) the
Variable Frequency Vibration Test,
as defined by Mil-Std-883 Method 2007. Both vibration tests are
similar in many aspects, although they are intended to uncover different
types of vibration-related failures.
The
vibration fatigue test
(Method 2005)
requires a testing apparatus that is capable of providing
sustained vibration
within the specified levels, and has the necessary optical and
electrical equipment needed to perform post-test measurements. In
this test, the sample is firmly fastened on the vibration platform with
its leads adequately secured. The sample is then subjected to a
simple harmonic vibration that has a constant amplitude having a peak
acceleration corresponding to the specified test condition.
The
vibration shall be performed for 32+/-8 hours minimum in each of the
X-, Y-, and Z- orientations, for a total of 96 hours minimum.
The test conditions applicable for this test are shown in Table 1.
Table 1. Test Conditions for the
Vibration Fatigue Test
per Mil-Std-883 Method 2005
Test Condition |
Stress Level (g) |
A |
20 |
B |
50 |
C |
70 |
The
variable frequency
vibration test (Method 2007)
requires a testing
apparatus that is capable of providing the required variable frequency
vibration at the specified levels, the necessary optical and
electrical equipment needed to perform post-test measurements.
In this test, the sample
is firmly fastened on the vibration platform with its leads adequately
secured. The sample is then subjected to a simple harmonic
vibration that has either a peak-to-peak amplitude of 0.06 inch +/- 10%
or a peak acceleration of the specified test condition.
The vibration frequency
shall then be varied approximately logarithmically between 20 and 2,000
Hz. The entire frequency range of 20-2000 Hz and the return to 20
Hz must be traversed in not less than 4 minutes. This cycle shall be
performed 4 times in each of the orientations X, Y, and Z, for a total
of 12 times.
The
test conditions applicable for this test are shown in Table 2.
Table 2. Test Conditions for the
Variable Frequency Vibration Test
per Mil-Std-883 Method 2007
Test Condition |
Stress Level (g) |
A |
20 |
B |
50 |
C |
70 |
After either vibration test has been completed,
external
visual inspection
of the case, leads, and seals shall be performed at 10 X to 20 X. The
marking shall also be inspected with or without magnification, but with
the magnification no greater than 3 X. An illegible mark and/or any
evidence of damage to the case, leads, or seals after the stress test
shall be considered a
failure.
Additional specified measurements may also be done after vibration testing.
Reference:
Mil-Std-883 Method 2005; Mil-Std-883 Method 2007
Reliability
Tests:
Autoclave
Test or PCT; Temperature
Cycling; Thermal
Shock;
THB;
HAST;
HTOL;
LTOL;
HTS; Solder
Heat Resistance Test (SHRT);
Other
Reliability Tests
See Also:
Reliability
Engineering;
Reliability Modeling; Qualification
Process; Failure
Analysis;
Package Failures; Die
Failures
HOME
Copyright
©
2001-2005
www.EESemi.com.
All Rights Reserved.